Recall Z-0180-2016— Encore Medical, Lp

Class II · Terminated

Class II FDA medical device recall by Encore Medical, Lp, initiated 2015-09-14, terminated 2016-11-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Reverse Shoulder Prosthesis Stem Impaction Fixture REF 804-03-053 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.. Reason: During the assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating..

Recalling firm
Encore Medical, Lp
Classification
Class II
Status
Terminated
Initiated
2015-09-14
Terminated
2016-11-04
Firm location
Austin, TX, United States

Product description

Reverse Shoulder Prosthesis Stem Impaction Fixture REF 804-03-053 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.

Reason for recall

During the assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Data sourced from openFDA. This site is unofficial and independent of the FDA.