Recall Z-0176-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-05-23, terminated 2018-03-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Dimension Vista Protein 1 Control H, PROT 1 CON H, Siemens Material Number (SMN) 10445921, Product Code KC717; IVD.. Reason: Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER)..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-05-23
Terminated
2018-03-30
Firm location
Newark, DE, United States

Product description

Dimension Vista Protein 1 Control H, PROT 1 CON H, Siemens Material Number (SMN) 10445921, Product Code KC717; IVD.

Reason for recall

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).