Recall Z-0171-2023— Becton, Dickinson and Company, BD Biosciences

Class II · Ongoing

Class II FDA medical device recall by Becton, Dickinson and Company, BD Biosciences, initiated 2022-10-03. It names one FDA 510(k) clearance: K090967. Product: BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.. Reason: Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier..

Recalling firm
Becton, Dickinson and Company, BD Biosciences
Classification
Class II
Status
Ongoing
Initiated
2022-10-03
Firm location
San Jose, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.

Reason for recall

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.