Recall Z-0166-2027— Howmedica Osteonics Corp.

Class II · Ongoing

Class II FDA medical device recall by Howmedica Osteonics Corp., initiated 2026-09-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization..

Recalling firm
Howmedica Osteonics Corp.
Classification
Class II
Status
Ongoing
Initiated
2026-09-16
Firm location
Mahwah, NJ, United States

Product description

Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.

Reason for recall

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Data sourced from openFDA. This site is unofficial and independent of the FDA.