Class II FDA medical device recall by Community Products, LLC, initiated 2019-03-21, terminated 2020-06-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Rifton E-Pacer, SKU K660, powered lift device. Reason: A design change to the load-bearing quick release buckle renders the springs in the latch mechanism susceptible to fatigue failure after repeated use..

Recalling firm
Community Products, LLC
Classification
Class II
Status
Terminated
Initiated
2019-03-21
Terminated
2020-06-15
Firm location
Rifton, NY, United States

Product description

Rifton E-Pacer, SKU K660, powered lift device

Reason for recall

A design change to the load-bearing quick release buckle renders the springs in the latch mechanism susceptible to fatigue failure after repeated use.