Recall Z-0165-2020— Community Products, LLC
Class II · Terminated
Class II FDA medical device recall by Community Products, LLC, initiated 2019-03-21, terminated 2020-06-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Rifton TRAM Low Base, SKU K320, powered lift device. Reason: A design change to the load-bearing quick release buckle renders the springs in the latch mechanism susceptible to fatigue failure after repeated use..
- Recalling firm
- Community Products, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-03-21
- Terminated
- 2020-06-15
- Firm location
- Rifton, NY, United States
Product description
Rifton TRAM Low Base, SKU K320, powered lift device
Reason for recall
A design change to the load-bearing quick release buckle renders the springs in the latch mechanism susceptible to fatigue failure after repeated use.
Data sourced from openFDA. This site is unofficial and independent of the FDA.