Recall Z-0160-2020— Siemens Medical Solutions USA, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2019-07-23, terminated 2020-05-06. It names one FDA 510(k) clearance: K173408. Product: Mammomat Revelation with software version VC10 and Biopsy Option Intended Use: For mammography exams, screening, diagnostics, biopsies and dual energy procedures. Reason: An error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare cases.
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-07-23
- Terminated
- 2020-05-06
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Mammomat Revelation with software version VC10 and Biopsy Option Intended Use: For mammography exams, screening, diagnostics, biopsies and dual energy procedures
Reason for recall
An error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare cases