Recall Z-0160-2018— Zimmer Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2014-11-06, terminated 2020-07-13. It names 2 FDA 510(k) clearances: K101791, K092239. Product: ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.. Reason: Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended)..

Recalling firm
Zimmer Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-11-06
Terminated
2020-07-13
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.

Reason for recall

Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).