Recall Z-0146-2015— Roche Diagnostics Operations, Inc.
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2014-09-25, terminated 2018-01-10. It names one FDA 510(k) clearance: K111353. Product: ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiple patient use. It is indicated for the treatment of insulin requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician.. Reason: The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the pump prompts customers to confirm the time and date, if the user does not recognize the time and date have returned to default, a shift of their basal rate time block could occur..
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-09-25
- Terminated
- 2018-01-10
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiple patient use. It is indicated for the treatment of insulin requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician.
Reason for recall
The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the pump prompts customers to confirm the time and date, if the user does not recognize the time and date have returned to default, a shift of their basal rate time block could occur.
Data sourced from openFDA. This site is unofficial and independent of the FDA.