Recall Z-0146-2014— Intuitive Surgical, Inc.

Class III · Terminated

Class III FDA medical device recall by Intuitive Surgical, Inc., initiated 2013-10-21, terminated 2014-09-04. It names one FDA 510(k) clearance: K081137. Product: da Vinci Vision System Cart, a component of the da Vinci Si Surgical System IS3000; Y1903 Illuminator (Illuminator + Lamp Module); Y1902 Replacement Lamp Modules. Intuitive Surgical, Sunnyvale, CA. As a component of the IS3000 da Vinci Si Surgical System, the 3DHD Vision System provides a high resolution image for the surgeon (at the 3D viewer) and the patient-side assistant (at the touchscreen). The 3DHD endoscope assembly (endoscope and camera head) can be used manually (that is, handheld, as with a traditional endoscopy cart) or can be mounted on the camera arm to assist with preoperative exploration. Also integrated in the Vision System Cart is the Illuminator. The illuminator has a single light source attached to the endoscopy assembly by the light guide cable. It provides illumination inside the body for vision of the surgical field. The Illuminator monitors the life left on the lamp module and notifies the customer when replacement is recommended.. Reason: Intuitive has identified that the metalized coating on the IS3000 HD lamp module may be incompatible with the illuminator control board, leading to an erroneous red Preventive Maintenance Message that shows up on monitors..

Recalling firm
Intuitive Surgical, Inc.
Classification
Class III
Status
Terminated
Initiated
2013-10-21
Terminated
2014-09-04
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

da Vinci Vision System Cart, a component of the da Vinci Si Surgical System IS3000; Y1903 Illuminator (Illuminator + Lamp Module); Y1902 Replacement Lamp Modules. Intuitive Surgical, Sunnyvale, CA. As a component of the IS3000 da Vinci Si Surgical System, the 3DHD Vision System provides a high resolution image for the surgeon (at the 3D viewer) and the patient-side assistant (at the touchscreen). The 3DHD endoscope assembly (endoscope and camera head) can be used manually (that is, handheld, as with a traditional endoscopy cart) or can be mounted on the camera arm to assist with preoperative exploration. Also integrated in the Vision System Cart is the Illuminator. The illuminator has a single light source attached to the endoscopy assembly by the light guide cable. It provides illumination inside the body for vision of the surgical field. The Illuminator monitors the life left on the lamp module and notifies the customer when replacement is recommended.

Reason for recall

Intuitive has identified that the metalized coating on the IS3000 HD lamp module may be incompatible with the illuminator control board, leading to an erroneous red Preventive Maintenance Message that shows up on monitors.