Recall Z-0136-2014— Bayer Healthcare Llc
Class II · Terminated
Class II FDA medical device recall by Bayer Healthcare Llc, initiated 2013-08-26, terminated 2014-01-14. It names one FDA 510(k) clearance: K090413. Product: Bayer brand A1CNow+ MULTI-TEST A1C SYSTEM TEST KITS (PART NUMBER 3021/08842610); Hemoglobin A1c Monitor; Glycosylated Hemoglobin Assay; Product is manufactured and distributed by Bayer HealthCare LLC, Sunnyvale, CA The A1CNow+ test provides quantitative measurement of the percent of glycated hemoglobin (%HbA1c or %A1C) levels in capillary (fingerstick) or venous whole blood samples. The test is for professional use to monitor glycemic control in people with diabetes.. Reason: Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release..
- Recalling firm
- Bayer Healthcare Llc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-08-26
- Terminated
- 2014-01-14
- Firm location
- Sunnyvale, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Bayer brand A1CNow+ MULTI-TEST A1C SYSTEM TEST KITS (PART NUMBER 3021/08842610); Hemoglobin A1c Monitor; Glycosylated Hemoglobin Assay; Product is manufactured and distributed by Bayer HealthCare LLC, Sunnyvale, CA The A1CNow+ test provides quantitative measurement of the percent of glycated hemoglobin (%HbA1c or %A1C) levels in capillary (fingerstick) or venous whole blood samples. The test is for professional use to monitor glycemic control in people with diabetes.
Reason for recall
Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.