Recall Z-0134-2016— Synthes (USA) Products LLC

Class II · Terminated

Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2015-09-04, terminated 2016-08-26. It names one FDA 510(k) clearance: K020517. Product: 5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm, 5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Unit Rod 370mm, 5.0mm Unit Rod 390mm, 5.0mm Unit Rod 410mm, 5.0mm Unit Rod 430mm, 5.0mm Unit Rod 450mm; Orthosis, Spinal, Pedicle fixation Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.. Reason: This product was produced using a finishing process not identified as part of the manufacturing specification. The process used with the lots subject to this Recall was a bead blast process. (Bead Blasting vs. Shot Peened)..

Recalling firm
Synthes (USA) Products LLC
Classification
Class II
Status
Terminated
Initiated
2015-09-04
Terminated
2016-08-26
Firm location
West Chester, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

5.0mm Unit Rod 270mm, 5.0mm Unit Rod 290mm, 5.0mm Unit Rod 310mm, 5.0mm Unit Rod 330mm, 5.0mm Unit Rod 350mm, 5.0mm Unit Rod 370mm, 5.0mm Unit Rod 390mm, 5.0mm Unit Rod 410mm, 5.0mm Unit Rod 430mm, 5.0mm Unit Rod 450mm; Orthosis, Spinal, Pedicle fixation Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.

Reason for recall

This product was produced using a finishing process not identified as part of the manufacturing specification. The process used with the lots subject to this Recall was a bead blast process. (Bead Blasting vs. Shot Peened).

Data sourced from openFDA. This site is unofficial and independent of the FDA.