Recall Z-0133-2027— Ventec Life Systems, Inc.
Class I · Ongoing
Class I FDA medical device recall by Ventec Life Systems, Inc., initiated 2026-09-17. FDA's recall record names no specific 510(k) clearance for this event. Product: V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.. Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy..
- Recalling firm
- Ventec Life Systems, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2026-09-17
- Firm location
- Bothell, WA, United States
Product description
V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.
Reason for recall
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.