Recall Z-0132-2021— Maquet Cardiovascular Us Sales, Llc

Class II · Terminated

Class II FDA medical device recall by Maquet Cardiovascular Us Sales, Llc, initiated 2020-09-24, terminated 2023-02-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143301F0 (US) Part 143301B0 (OUS). Reason: Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components.

Recalling firm
Maquet Cardiovascular Us Sales, Llc
Classification
Class II
Status
Terminated
Initiated
2020-09-24
Terminated
2023-02-28
Firm location
Wayne, NJ, United States

Product description

Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143301F0 (US) Part 143301B0 (OUS)

Reason for recall

Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components

Data sourced from openFDA. This site is unofficial and independent of the FDA.