Recall Z-0121-2020— Centurion Medical Products Corporation
Class II · Terminated
Class II FDA medical device recall by Centurion Medical Products Corporation, initiated 2019-08-28, terminated 2020-07-23. It names one FDA 510(k) clearance: K890897. Product: Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR. Reason: A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel..
- Recalling firm
- Centurion Medical Products Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-08-28
- Terminated
- 2020-07-23
- Firm location
- Williamston, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR
Reason for recall
A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
Data sourced from openFDA. This site is unofficial and independent of the FDA.