Recall Z-0119-2020— Horiba Instruments Incorporated

Class II · Terminated

Class II FDA medical device recall by Horiba Instruments Incorporated, initiated 2018-10-30, terminated 2020-06-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944. Reason: Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory)..

Recalling firm
Horiba Instruments Incorporated
Classification
Class II
Status
Terminated
Initiated
2018-10-30
Terminated
2020-06-16
Firm location
Irvine, CA, United States

Product description

Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944

Reason for recall

Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).