Recall Z-0117-2023— Abbott Vascular

Class II · Ongoing

Class II FDA medical device recall by Abbott Vascular, initiated 2022-09-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT, MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT, TriClip G4 NT Delivery System CLIN/TCDS0303-NT, TriClip G4 NTW Delivery System CLIN/TCDS0303-NTW, TriClip G4 XT Delivery System CLIN/TCDS0303-XT, TriClip G4 XTW Delivery System CLIN/TCDS0303-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT. For cardiac procedures.. Reason: An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury..

Recalling firm
Abbott Vascular
Classification
Class II
Status
Ongoing
Initiated
2022-09-08
Firm location
Santa Clara, CA, United States

Product description

Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT, MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT, TriClip G4 NT Delivery System CLIN/TCDS0303-NT, TriClip G4 NTW Delivery System CLIN/TCDS0303-NTW, TriClip G4 XT Delivery System CLIN/TCDS0303-XT, TriClip G4 XTW Delivery System CLIN/TCDS0303-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT. For cardiac procedures.

Reason for recall

An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

Data sourced from openFDA. This site is unofficial and independent of the FDA.