Recall Z-0116-2022— Elekta, Inc.
Class II · Terminated
Class II FDA medical device recall by Elekta, Inc., initiated 2021-09-17, terminated 2024-09-30. It names one FDA 510(k) clearance: K210500. Product: Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD. Reason: There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter..
- Recalling firm
- Elekta, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-09-17
- Terminated
- 2024-09-30
- Firm location
- Atlanta, GA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD
Reason for recall
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.
Data sourced from openFDA. This site is unofficial and independent of the FDA.