Recall Z-0115-2022— ELEKTA SOLUTIONS AB
Class II · Terminated
Class II FDA medical device recall by ELEKTA SOLUTIONS AB, initiated 2021-09-28, terminated 2024-06-14. It names one FDA 510(k) clearance: K210500. Product: Elekta Medical Linear Accelerator (EMLA): Elekta Harmony Pro, REF 1555143; Elekta Infinity, REF 1552091, and Elekta Versa HD, REF 1552093.. Reason: There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter..
- Recalling firm
- ELEKTA SOLUTIONS AB
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-09-28
- Terminated
- 2024-06-14
- Firm location
- Stockholm, N/A, Sweden
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Elekta Medical Linear Accelerator (EMLA): Elekta Harmony Pro, REF 1555143; Elekta Infinity, REF 1552091, and Elekta Versa HD, REF 1552093.
Reason for recall
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.