Recall Z-0114-2014— Synthes USA HQ, Inc.

Class II · Terminated

Class II FDA medical device recall by Synthes USA HQ, Inc., initiated 2013-08-13, terminated 2016-08-18. It names one FDA 510(k) clearance: K110789. Product: Synthes Application Instrument Sternal Zip Flex System intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion. Reason: When using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant..

Recalling firm
Synthes USA HQ, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-08-13
Terminated
2016-08-18
Firm location
West Chester, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Synthes Application Instrument Sternal Zip Flex System intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion

Reason for recall

When using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant.