Recall Z-0113-2020— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2019-09-03, terminated 2021-02-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Achilles Express Bone Sonometer. Reason: Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs are required for protective earth connection..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2019-09-03
Terminated
2021-02-16
Firm location
Waukesha, WI, United States

Product description

Achilles Express Bone Sonometer

Reason for recall

Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs are required for protective earth connection.