Recall Z-0110-2019— GE Medical Systems, LLC

Class II · Terminated

Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2018-06-07, terminated 2020-06-18. It names 6 FDA 510(k) clearances: K052293, K041476, K040444 and 3 more. Product: GE Healthcare MRI Systems, Software Version 11.x, sold under the following product names: 1.5T CXK4 LCC MAGNET,1.5T EXCITE, 1.5T EXCITE E/S, 1.5T EXCITE GS, 1.5T EXCITE HYBRID, 1.5T EXCITE TWIN, 1.5T EXCITE UPGRAD, 1.5T HD ES/HS TR W/EXCITE, 1.5T MR MOBILE (SERIAL # R350), 1.5T SIGNA ECHOSPEED, 1.5T SIGNA EXCITE, 1.5T SIGNA EXCITE MRI, 1.5T SIGNA VIBRANT 16CH FIXED, 3.0T MRI TWINSPEED 8CH, 3.0T SIGNA EXCITE FIXED, 3.0T SIGNA EXCITE UPGRADE, 3T EXCITE, EX ES+8CH USED EXCITE, EXCITE 1.5T, EXCITE 3.0T, EXCITE TWIN, GOLDSEAL 1.5T EXCITE 4CH, GOLDSEAL 1.5T EXCITE 8CH, GOLDSEAL 1.5T EXCITE REL GOLDSEAL EXCITE 1.5T, GOLDSEAL EXCITE 1.5T 8CH, GS 1.5 EX ES+ 4CH NEW, GS 1.5 EX ES+4CH USED, GS 1.5 EX ES+8CH NEW, GS 1.5T EXCITE MRI, GS CLASS 1.5T ES+ EX 4CHN, GS SIGNA EXCITE II 1.5T, GS SIGNA MR/I 1.5T, LCC EXCITE 1.5T, MR 1.5 EX ES 8CH NEW, MR 1.5 EX ES+ 4CH, MR 1.5 EX ES+4CH MOB, MR 1.5 EX ES+8CH, MR 1.5EX8CH, MR 1.5EX8CH USE -HUNG, MR 1.5T EX8CH, MR 1.5T EXCITE 4CH, MR 1.5T EXCITE EXPERT MR, MR 1.5T EXCITE GS, MR 1.5T EXCITE GS MR, MR 1.5T GS ES+ EXCITE 4 CHANNE, MR 1.5T SIGNA EXCITE 8 CH GOLD SEAL, MR 1.5T SIGNA EXCITE II 4CH, MR 3.0T TWINSPEED, MR 3.0T SIGNA EXCI, MR 3.0T SIGNA EXCITE, MR 3.0T SIGNA EXCITE FIXED, MR EXCITE, MR EXCITE 1.5T, MR EXCITE 3, MR EXCITE ECHOSPEE, MR EXCITE ECHOSPEE UPG, MR EXCITE ECHOSPEED, MR EXCITE HISPEED MR EXCITE TWINSPEE, MR EXCITE TWINSPEED, MR GOLD SEAL 1.5 EX ES+8CH NEW, MR GOLDSEAL 1.5 ES PL EX 4 CH, MR GOLDSEAL 1.5 EX, MR GOLDSEAL 1.5 EX, MR GOLDSEAL 1.5 EX 4CH, MR GOLDSEAL 1.5 EX 8CH, MR GOLDSEAL 1.5 EX ES+ 4CH, MR GOLDSEAL 1.5 EX ES+ 4CH NEW, MR GOLDSEAL 1.5 EX ES+4CH, MR GOLDSEAL 1.5 EX ES+4CH USD, MR GOLDSEAL 1.5 EX ES+4CH USE, MR GOLDSEAL 1.5 EX ES+4CH USED, R GOLDSEAL 1.5 EX ES+8CH, MR GOLDSEAL 1.5 EX ES+8CH NEW, MR GOLDSEAL 1.5 EX ES+8CH USED, MR GOLDSEAL 1.5 EXCITE S+ 4CH, MR GOLDSEAL 1.5 FIXED ACTIVE, MR GOLDSEAL 1.5EX4CH NEW -HUNG, MR GOLDSEAL 3.0T EXCITE HD FIX, MR GS 1.5 EX, MR GS 1.5 EX ES+8C, MR GS 1.5 EX ES+8CH, MR SIGNA 1.5 EXCIT, MR SIGNA 1.5 EXCITE EXPERT, MR SIGNA 1.5T EXCITE, MR SIGNA 1.5T EXCITE 4CH MR SIGNA 1.5T EXCITE 8CH, MR SIGNA 1.5T EXCITE 8CH MOBIL, MR SIGNA 1.5T EXCITE EXPERT, MR SIGNA 1.5T EXCITE II, MR SIGNA 15T EXCITE 8CH, MR SIGNA 15T EXCITE 8CH MOBIL, MR SIGNA EXCITE 1.5 T GS (4 CHANNEL, MR SIGNA MR/I ECHOS W/EXCITE, MR SIGNA MR/I ECHOSPD W/ EXCITE, MR SIGNA MR/I ECHOSPD W/ EXCITE , MR SIGNA MR/I ECHOSPEED W/ EXCITE, MR SIGNA MR/I HISPEED W/ EXCITE MR SIGNA MR/I TWISP TR W/EXCIT, MR TWINSPEED 1.5T W/ EXCITE, MR UPG 1.5 SIGNA LX TO EXCITE, MR UPG 3.0T94 ECLIPSE TO EXCIT, MR/I ECHOSPD W/ EXCITE, MR/I ECHOSPEED W/ EXCITE, MR/I SIGNA ECHOSPD W/ EXCITE, MRI EXCITE, MRI EXCITE 1.5T FIXED, MRI SIGNA EXCITE 1.5T, PREOWN 1.5T EXITE, SIGNA 1.5 EXCITE, SIGNA 1.5T , CHOSPEED W EXCITE, SIGNA 1.5T ECHOSPEED W/EXCITE, SIGNA 1.5T ES EXCITE, SIGNA 1.5T EXCITE, SIGNA 1.5T EXCITE 2, SIGNA 1.5T EXCITE EXPERT, SIGNA 1.5T EXCITE II, SIGNA 1.5T SMART SPEED, /EXCITE, SIGNA 1.5T W/ EXCITE, SIGNA 3.0T E, SIGNA 3.0T SIGNA EXCITE FIXED, SIGNA ECHOSPEED 1.5T FIXED, SIGNA ECHOSPEED EXCITE 1.5T, SIGNA ECHOSPEED LX 1.5T FIXED SIGNA ECHOSPEED PLUS + EXCITE, SIGNA EXCITE, SIGNA EXCITE 1.5 SIGNA EXCITE 1.5 T HISPEED, SIGNA EXCITE 1.5T, SIGNA EXCITE 1.5T RELEASE 12.X/MR, SIGNA EXCITE 1.5T SMART SPEED, SIGNA EXCITE 11X, SIGNA EXCITE 3.0T, SIGNA EXCITE ECHOSPEED 1,5 T, SIGNA EXCITE ECHOSPEEED PLUS 1.5 T, SIGNA EXCITE EXPERT 1.5T, SIGNA EXCITE GS 1.5T 4 CHANNEL,SIGNA EXCITE II, SIGNA HISPD LX K4 1.5T FIXED, SIGNA HORIZON LX 1.5T FIXED, SIGNA LX 1.5 EXCITE II., SIGNA LX 1.5T EXCITE, SIGNA LX 1.5T EXCITE II, SIGNA LX1.5T HISPEED, SIGNA MR 1.5 ES EXCITE, SIGNA MR/I 1.5T EXCITE, SIGNA MR/I ECHO W/E,SIGNA MR/I ECHOESPEED W/EXCITE, SIGNA MR/I ECHOSPD, SIGNA MR/I ECHOSPD 1.5T FIXED, SIGNA MR/I ECHOSPD W/ EXCITE 1.5, SIGNA MR/I ECHOSPEED W/ EXCITE, SIGNA MR/I ECHOSPEED W/ EXCITE 1.5T, SIGNA MR/I ECHOSPEED W/EXCITE, SIGNA MR/I ECHOSPEED WITH EXCI, SIGNA MR/I ECHOSPEED WITH EXCITE SIGNA, MR/I. Reason: The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch..

Recalling firm
GE Medical Systems, LLC
Classification
Class II
Status
Terminated
Initiated
2018-06-07
Terminated
2020-06-18
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

6 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GE Healthcare MRI Systems, Software Version 11.x, sold under the following product names: 1.5T CXK4 LCC MAGNET,1.5T EXCITE, 1.5T EXCITE E/S, 1.5T EXCITE GS, 1.5T EXCITE HYBRID, 1.5T EXCITE TWIN, 1.5T EXCITE UPGRAD, 1.5T HD ES/HS TR W/EXCITE, 1.5T MR MOBILE (SERIAL # R350), 1.5T SIGNA ECHOSPEED, 1.5T SIGNA EXCITE, 1.5T SIGNA EXCITE MRI, 1.5T SIGNA VIBRANT 16CH FIXED, 3.0T MRI TWINSPEED 8CH, 3.0T SIGNA EXCITE FIXED, 3.0T SIGNA EXCITE UPGRADE, 3T EXCITE, EX ES+8CH USED EXCITE, EXCITE 1.5T, EXCITE 3.0T, EXCITE TWIN, GOLDSEAL 1.5T EXCITE 4CH, GOLDSEAL 1.5T EXCITE 8CH, GOLDSEAL 1.5T EXCITE REL GOLDSEAL EXCITE 1.5T, GOLDSEAL EXCITE 1.5T 8CH, GS 1.5 EX ES+ 4CH NEW, GS 1.5 EX ES+4CH USED, GS 1.5 EX ES+8CH NEW, GS 1.5T EXCITE MRI, GS CLASS 1.5T ES+ EX 4CHN, GS SIGNA EXCITE II 1.5T, GS SIGNA MR/I 1.5T, LCC EXCITE 1.5T, MR 1.5 EX ES 8CH NEW, MR 1.5 EX ES+ 4CH, MR 1.5 EX ES+4CH MOB, MR 1.5 EX ES+8CH, MR 1.5EX8CH, MR 1.5EX8CH USE -HUNG, MR 1.5T EX8CH, MR 1.5T EXCITE 4CH, MR 1.5T EXCITE EXPERT MR, MR 1.5T EXCITE GS, MR 1.5T EXCITE GS MR, MR 1.5T GS ES+ EXCITE 4 CHANNE, MR 1.5T SIGNA EXCITE 8 CH GOLD SEAL, MR 1.5T SIGNA EXCITE II 4CH, MR 3.0T TWINSPEED, MR 3.0T SIGNA EXCI, MR 3.0T SIGNA EXCITE, MR 3.0T SIGNA EXCITE FIXED, MR EXCITE, MR EXCITE 1.5T, MR EXCITE 3, MR EXCITE ECHOSPEE, MR EXCITE ECHOSPEE UPG, MR EXCITE ECHOSPEED, MR EXCITE HISPEED MR EXCITE TWINSPEE, MR EXCITE TWINSPEED, MR GOLD SEAL 1.5 EX ES+8CH NEW, MR GOLDSEAL 1.5 ES PL EX 4 CH, MR GOLDSEAL 1.5 EX, MR GOLDSEAL 1.5 EX, MR GOLDSEAL 1.5 EX 4CH, MR GOLDSEAL 1.5 EX 8CH, MR GOLDSEAL 1.5 EX ES+ 4CH, MR GOLDSEAL 1.5 EX ES+ 4CH NEW, MR GOLDSEAL 1.5 EX ES+4CH, MR GOLDSEAL 1.5 EX ES+4CH USD, MR GOLDSEAL 1.5 EX ES+4CH USE, MR GOLDSEAL 1.5 EX ES+4CH USED, R GOLDSEAL 1.5 EX ES+8CH, MR GOLDSEAL 1.5 EX ES+8CH NEW, MR GOLDSEAL 1.5 EX ES+8CH USED, MR GOLDSEAL 1.5 EXCITE S+ 4CH, MR GOLDSEAL 1.5 FIXED ACTIVE, MR GOLDSEAL 1.5EX4CH NEW -HUNG, MR GOLDSEAL 3.0T EXCITE HD FIX, MR GS 1.5 EX, MR GS 1.5 EX ES+8C, MR GS 1.5 EX ES+8CH, MR SIGNA 1.5 EXCIT, MR SIGNA 1.5 EXCITE EXPERT, MR SIGNA 1.5T EXCITE, MR SIGNA 1.5T EXCITE 4CH MR SIGNA 1.5T EXCITE 8CH, MR SIGNA 1.5T EXCITE 8CH MOBIL, MR SIGNA 1.5T EXCITE EXPERT, MR SIGNA 1.5T EXCITE II, MR SIGNA 15T EXCITE 8CH, MR SIGNA 15T EXCITE 8CH MOBIL, MR SIGNA EXCITE 1.5 T GS (4 CHANNEL, MR SIGNA MR/I ECHOS W/EXCITE, MR SIGNA MR/I ECHOSPD W/ EXCITE, MR SIGNA MR/I ECHOSPD W/ EXCITE , MR SIGNA MR/I ECHOSPEED W/ EXCITE, MR SIGNA MR/I HISPEED W/ EXCITE MR SIGNA MR/I TWISP TR W/EXCIT, MR TWINSPEED 1.5T W/ EXCITE, MR UPG 1.5 SIGNA LX TO EXCITE, MR UPG 3.0T94 ECLIPSE TO EXCIT, MR/I ECHOSPD W/ EXCITE, MR/I ECHOSPEED W/ EXCITE, MR/I SIGNA ECHOSPD W/ EXCITE, MRI EXCITE, MRI EXCITE 1.5T FIXED, MRI SIGNA EXCITE 1.5T, PREOWN 1.5T EXITE, SIGNA 1.5 EXCITE, SIGNA 1.5T , CHOSPEED W EXCITE, SIGNA 1.5T ECHOSPEED W/EXCITE, SIGNA 1.5T ES EXCITE, SIGNA 1.5T EXCITE, SIGNA 1.5T EXCITE 2, SIGNA 1.5T EXCITE EXPERT, SIGNA 1.5T EXCITE II, SIGNA 1.5T SMART SPEED, /EXCITE, SIGNA 1.5T W/ EXCITE, SIGNA 3.0T E, SIGNA 3.0T SIGNA EXCITE FIXED, SIGNA ECHOSPEED 1.5T FIXED, SIGNA ECHOSPEED EXCITE 1.5T, SIGNA ECHOSPEED LX 1.5T FIXED SIGNA ECHOSPEED PLUS + EXCITE, SIGNA EXCITE, SIGNA EXCITE 1.5 SIGNA EXCITE 1.5 T HISPEED, SIGNA EXCITE 1.5T, SIGNA EXCITE 1.5T RELEASE 12.X/MR, SIGNA EXCITE 1.5T SMART SPEED, SIGNA EXCITE 11X, SIGNA EXCITE 3.0T, SIGNA EXCITE ECHOSPEED 1,5 T, SIGNA EXCITE ECHOSPEEED PLUS 1.5 T, SIGNA EXCITE EXPERT 1.5T, SIGNA EXCITE GS 1.5T 4 CHANNEL,SIGNA EXCITE II, SIGNA HISPD LX K4 1.5T FIXED, SIGNA HORIZON LX 1.5T FIXED, SIGNA LX 1.5 EXCITE II., SIGNA LX 1.5T EXCITE, SIGNA LX 1.5T EXCITE II, SIGNA LX1.5T HISPEED, SIGNA MR 1.5 ES EXCITE, SIGNA MR/I 1.5T EXCITE, SIGNA MR/I ECHO W/E,SIGNA MR/I ECHOESPEED W/EXCITE, SIGNA MR/I ECHOSPD, SIGNA MR/I ECHOSPD 1.5T FIXED, SIGNA MR/I ECHOSPD W/ EXCITE 1.5, SIGNA MR/I ECHOSPEED W/ EXCITE, SIGNA MR/I ECHOSPEED W/ EXCITE 1.5T, SIGNA MR/I ECHOSPEED W/EXCITE, SIGNA MR/I ECHOSPEED WITH EXCI, SIGNA MR/I ECHOSPEED WITH EXCITE SIGNA, MR/I

Reason for recall

The latest on-site software version may not have been reinstalled at some sites after service activities were performed resulting in images possibly flipped left/right and/or there can be patient data mismatch.