Recall Z-0109-2017— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-07-15, terminated 2017-06-05. FDA's recall record names no specific 510(k) clearance for this event. Product: cobas p 312 Pre-Analytical System calculator/data processing module, for clinical use Product Usage: The cobas p471 system is an automatic centrifuge module for centrifuging primary sample tubes, which will then be transported to the output module.. Reason: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-07-15
Terminated
2017-06-05
Firm location
Indianapolis, IN, United States

Product description

cobas p 312 Pre-Analytical System calculator/data processing module, for clinical use Product Usage: The cobas p471 system is an automatic centrifuge module for centrifuging primary sample tubes, which will then be transported to the output module.

Reason for recall

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.