Recall Z-0108-2027— Canon Medical System, USA, INC.

Class II · Ongoing

Class II FDA medical device recall by Canon Medical System, USA, INC., initiated 2026-08-26. It names one FDA 510(k) clearance: K210900. Product: Alphenix INFX-8000C/B5. Reason: Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error.

Recalling firm
Canon Medical System, USA, INC.
Classification
Class II
Status
Ongoing
Initiated
2026-08-26
Firm location
Tustin, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Alphenix INFX-8000C/B5

Reason for recall

Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error

Data sourced from openFDA. This site is unofficial and independent of the FDA.