Recall Z-0107-2022— Smiths Medical ASD Inc.
Class I · Terminated
Class I FDA medical device recall by Smiths Medical ASD Inc., initiated 2020-03-24, terminated 2023-12-22. It names one FDA 510(k) clearance: K111275. Product: CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L. Reason: Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician..
- Recalling firm
- Smiths Medical ASD Inc.
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2020-03-24
- Terminated
- 2023-12-22
- Firm location
- Minneapolis, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L
Reason for recall
Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.