Recall Z-0105-2021— Cook Inc.

Class II · Terminated

Class II FDA medical device recall by Cook Inc., initiated 2020-08-31, terminated 2021-07-01. It names one FDA 510(k) clearance: K181735. Product: Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE-coated stainless steel; a 5, 6, and 7 French, 10 cm radiopaque dilator; a 6 French, 7 cm radiopaque pigtail catheter; a retention disc with pull tie; and a 10 French, 10 cm clear, polyvinylchloride connecting tube with one-way stopcock.. Reason: The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation..

Recalling firm
Cook Inc.
Classification
Class II
Status
Terminated
Initiated
2020-08-31
Terminated
2021-07-01
Firm location
Bloomington, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE-coated stainless steel; a 5, 6, and 7 French, 10 cm radiopaque dilator; a 6 French, 7 cm radiopaque pigtail catheter; a retention disc with pull tie; and a 10 French, 10 cm clear, polyvinylchloride connecting tube with one-way stopcock.

Reason for recall

The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.

Data sourced from openFDA. This site is unofficial and independent of the FDA.