Recall Z-0104-2017— Roche Diagnostics Operations, Inc.
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-07-15, terminated 2017-06-05. FDA's recall record names no specific 510(k) clearance for this event. Product: RSD 800A Aliquoting System calculator/data processing module, for clinical use Product Usage: The Roche/Hitachi MODULAR Analytics System is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests.. Reason: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process..
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-07-15
- Terminated
- 2017-06-05
- Firm location
- Indianapolis, IN, United States
Product description
RSD 800A Aliquoting System calculator/data processing module, for clinical use Product Usage: The Roche/Hitachi MODULAR Analytics System is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests.
Reason for recall
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Data sourced from openFDA. This site is unofficial and independent of the FDA.