Recall Z-0101-2023— Synthes (USA) Products LLC

Class II · Ongoing

Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2022-09-27. It names one FDA 510(k) clearance: K191463. Product: DePuy Synthes Hammertoe Continuous Compression Implants, Product No. 46.239.001. Reason: The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit..

Recalling firm
Synthes (USA) Products LLC
Classification
Class II
Status
Ongoing
Initiated
2022-09-27
Firm location
West Chester, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

DePuy Synthes Hammertoe Continuous Compression Implants, Product No. 46.239.001

Reason for recall

The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.