Recall Z-0101-2017— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-07-15, terminated 2017-06-05. FDA's recall record names no specific 510(k) clearance for this event. Product: cobas p 701 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The URISYS 2400 Analyzer is a fully automated urinalysis system intended for in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and erythrocytes, In addition, the analyzer determines specific gravity, color, and clarity. The URISYS 2400 is intended for professional use only.. Reason: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-07-15
Terminated
2017-06-05
Firm location
Indianapolis, IN, United States

Product description

cobas p 701 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The URISYS 2400 Analyzer is a fully automated urinalysis system intended for in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and erythrocytes, In addition, the analyzer determines specific gravity, color, and clarity. The URISYS 2400 is intended for professional use only.

Reason for recall

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Data sourced from openFDA. This site is unofficial and independent of the FDA.