Recall Z-0097-2017— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2016-07-15, terminated 2017-06-05. It names one FDA 510(k) clearance: K100853. Product: cobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical use Modular e602: Immunoassay Analyzer Product Usage: The VS II system is a stand-alone, computer-controlled instrument designed to aliquot patient sample from a primary tube to a barcoded secondary tube for processing on an off-line analyzer.. Reason: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-07-15
Terminated
2017-06-05
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

cobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical use Modular e602: Immunoassay Analyzer Product Usage: The VS II system is a stand-alone, computer-controlled instrument designed to aliquot patient sample from a primary tube to a barcoded secondary tube for processing on an off-line analyzer.

Reason for recall

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Data sourced from openFDA. This site is unofficial and independent of the FDA.