Recall Z-0095-2019— GE Healthcare Finland Oy

Class II · Ongoing

Class II FDA medical device recall by GE Healthcare Finland Oy, initiated 2018-09-04. It names one FDA 510(k) clearance: K132533. Product: GE Healthcare CARESCAPE Monitor B450 Product Usage: Intended uses of CARESCAPE B450 in 510(k) countries [UR_CRSCP-885]: The CARESCAPE Monitor B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B450 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B450 is indicated for monitoring of: -hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation, -Respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), and spirometry) -neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).. Reason: When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected..

Recalling firm
GE Healthcare Finland Oy
Classification
Class II
Status
Ongoing
Initiated
2018-09-04
Firm location
Helsinki, Finland

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GE Healthcare CARESCAPE Monitor B450 Product Usage: Intended uses of CARESCAPE B450 in 510(k) countries [UR_CRSCP-885]: The CARESCAPE Monitor B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B450 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B450 is indicated for monitoring of: -hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation, -Respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), and spirometry) -neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).

Reason for recall

When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.