Recall Z-0090-2025— LEICA BIOSYSTEMS NUSSLOCH GMBH

Class II · Ongoing

Class II FDA medical device recall by LEICA BIOSYSTEMS NUSSLOCH GMBH, initiated 2024-09-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Leica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 14903050S03; for freezing and sectioning tissue samples. Reason: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019..

Recalling firm
LEICA BIOSYSTEMS NUSSLOCH GMBH
Classification
Class II
Status
Ongoing
Initiated
2024-09-04
Firm location
Nussloch, Germany

Product description

Leica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 14903050S03; for freezing and sectioning tissue samples

Reason for recall

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Data sourced from openFDA. This site is unofficial and independent of the FDA.