Recall Z-0087-2023— Shanghai United Imaging Healthcare Co., Ltd.

Class II · Ongoing

Class II FDA medical device recall by Shanghai United Imaging Healthcare Co., Ltd., initiated 2022-05-30. It names 2 FDA 510(k) clearances: K193241, K182237. Product: Positron Emission Tomography and Computed Tomography System, Model: uMI 550. Reason: The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient..

Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Classification
Class II
Status
Ongoing
Initiated
2022-05-30
Firm location
Shanghai, N/A, China

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Reason for recall

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.