Recall Z-0086-2015— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2014-09-18, terminated 2015-02-24. It names one FDA 510(k) clearance: K122286. Product: Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.. Reason: The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2014-09-18
Terminated
2015-02-24
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Reason for recall

The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i