Recall Z-0085-2027— Stryker Sustainability Solutions

Class II · Ongoing

Class II FDA medical device recall by Stryker Sustainability Solutions, initiated 2026-08-18. It names one FDA 510(k) clearance: K030187. Product: BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007. Reason: Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could lead to particulate embolization, infection, thrombus or ganuloma formation from particulate matter..

Recalling firm
Stryker Sustainability Solutions
Classification
Class II
Status
Ongoing
Initiated
2026-08-18
Firm location
Tempe, AZ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007

Reason for recall

Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could lead to particulate embolization, infection, thrombus or ganuloma formation from particulate matter.

Data sourced from openFDA. This site is unofficial and independent of the FDA.