Recall Z-0084-2026— Boston Scientific Corporation
Class I · Ongoing
Class I FDA medical device recall by Boston Scientific Corporation, initiated 2025-08-20. FDA's recall record names no specific 510(k) clearance for this event. Product: ACCOLADE DR EL MRI (Model L331). Reason: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps)..
- Recalling firm
- Boston Scientific Corporation
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2025-08-20
- Firm location
- Saint Paul, MN, United States
Product description
ACCOLADE DR EL MRI (Model L331)
Reason for recall
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Data sourced from openFDA. This site is unofficial and independent of the FDA.