Recall Z-0083-2026— Boston Scientific Corporation

Class I · Ongoing

Class I FDA medical device recall by Boston Scientific Corporation, initiated 2025-08-20. FDA's recall record names no specific 510(k) clearance for this event. Product: ACCOLADE DR EL (Model L321). Reason: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps)..

Recalling firm
Boston Scientific Corporation
Classification
Class I
Status
Ongoing
Initiated
2025-08-20
Firm location
Saint Paul, MN, United States

Product description

ACCOLADE DR EL (Model L321)

Reason for recall

Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

Data sourced from openFDA. This site is unofficial and independent of the FDA.