Recall Z-0079-2027— Cardinal Health 200, LLC

Class II · Ongoing

Class II FDA medical device recall by Cardinal Health 200, LLC, initiated 2026-08-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G. Reason: Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation..

Recalling firm
Cardinal Health 200, LLC
Classification
Class II
Status
Ongoing
Initiated
2026-08-19
Firm location
Waukegan, IL, United States

Product description

Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G

Reason for recall

Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.

Data sourced from openFDA. This site is unofficial and independent of the FDA.