Recall Z-0076-2014— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2013-08-30, terminated 2014-09-25. It names one FDA 510(k) clearance: K033363. Product: Zimmer¿ Unicompartmental Knee Femoral Component. Indicated for patients with painful and/or disabling knee joint.. Reason: The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-08-30
Terminated
2014-09-25
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Zimmer¿ Unicompartmental Knee Femoral Component. Indicated for patients with painful and/or disabling knee joint.

Reason for recall

The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.

Data sourced from openFDA. This site is unofficial and independent of the FDA.