Recall Z-0073-2025— Merit Medical Systems, Inc.

Class II · Ongoing

Class II FDA medical device recall by Merit Medical Systems, Inc., initiated 2024-08-19. FDA's recall record names no specific 510(k) clearance for this event. Product: StabiliT TOUCH Syringe, REF: IN8VCF/B. Reason: Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection..

Recalling firm
Merit Medical Systems, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-08-19
Firm location
South Jordan, UT, United States

Product description

StabiliT TOUCH Syringe, REF: IN8VCF/B

Reason for recall

Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.

Data sourced from openFDA. This site is unofficial and independent of the FDA.