Recall Z-0072-2020— Becton, Dickinson and Company, BD Biosciences
Class II · Terminated
Class II FDA medical device recall by Becton, Dickinson and Company, BD Biosciences, initiated 2019-08-19, terminated 2024-04-12. It names one FDA 510(k) clearance: K170974. Product: BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 654587, when using BD Trucount Tubes. Reason: Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion..
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-08-19
- Terminated
- 2024-04-12
- Firm location
- San Jose, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 654587, when using BD Trucount Tubes
Reason for recall
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Data sourced from openFDA. This site is unofficial and independent of the FDA.