Recall Z-0069-2020— Becton, Dickinson and Company, BD Biosciences

Class II · Terminated

Class II FDA medical device recall by Becton, Dickinson and Company, BD Biosciences, initiated 2019-08-19, terminated 2024-04-12. It names one FDA 510(k) clearance: K170974. Product: BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes. Reason: Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion..

Recalling firm
Becton, Dickinson and Company, BD Biosciences
Classification
Class II
Status
Terminated
Initiated
2019-08-19
Terminated
2024-04-12
Firm location
San Jose, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes

Reason for recall

Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.

Data sourced from openFDA. This site is unofficial and independent of the FDA.