Recall Z-0064-2019— Leica Microsystems, Inc.

Class II · Terminated

Class II FDA medical device recall by Leica Microsystems, Inc., initiated 2018-06-13, terminated 2019-03-18. FDA's recall record names no specific 510(k) clearance for this event. Product: Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.. Reason: These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user..

Recalling firm
Leica Microsystems, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-06-13
Terminated
2019-03-18
Firm location
Buffalo Grove, IL, United States

Product description

Leica Biosystems HistoCore SPECTRA ST, Slide Stainer, Automated Product Usage: The intended use of this device is a stainer for laboratory applications and intended for the preparation of histological and cytological specimens.

Reason for recall

These devices have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.

Data sourced from openFDA. This site is unofficial and independent of the FDA.