Recall Z-0060-2015— GE Healthcare, LLC
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2014-09-19, terminated 2016-09-13. It names one FDA 510(k) clearance: K071073. Product: GE Healthcare Carescape Patient Data Module. Reason: GE Healthcare has recently become aware of a potential safety issue to the ECG calculations following a disconnect/reconnect cycle with the Patient Data Module, when used with the Carescape Bx50 monitors. If the PDM is quickly disconnected and reconnected (a less than 10 second cycle) from a Carescape Bx50 monitor following a Patient Discharge or during active monitoring, the ECG waveform an.
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-09-19
- Terminated
- 2016-09-13
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare Carescape Patient Data Module
Reason for recall
GE Healthcare has recently become aware of a potential safety issue to the ECG calculations following a disconnect/reconnect cycle with the Patient Data Module, when used with the Carescape Bx50 monitors. If the PDM is quickly disconnected and reconnected (a less than 10 second cycle) from a Carescape Bx50 monitor following a Patient Discharge or during active monitoring, the ECG waveform an