Recall Z-0057-2015— Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Biomet, Inc., initiated 2014-09-26, terminated 2015-07-14. FDA's recall record names no specific 510(k) clearance for this event. Product: Green Mamba Suture Passer. The Biomet Sports Medicine Mamba instruments are utilized to aid in passing suture through soft tissue. The Mamba Needle is designed for use with the Mamba Suture Passer. The needles are sterile, disposable, and for single-patient use only.. Reason: An investigation identified that high level friction may exist between the needle and Mamba suture passer instrument, causing the needle to break during use..

Recalling firm
Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-09-26
Terminated
2015-07-14
Firm location
Warsaw, IN, United States

Product description

Green Mamba Suture Passer. The Biomet Sports Medicine Mamba instruments are utilized to aid in passing suture through soft tissue. The Mamba Needle is designed for use with the Mamba Suture Passer. The needles are sterile, disposable, and for single-patient use only.

Reason for recall

An investigation identified that high level friction may exist between the needle and Mamba suture passer instrument, causing the needle to break during use.

Data sourced from openFDA. This site is unofficial and independent of the FDA.