Recall Z-0046-2013— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2012-07-24, terminated 2013-10-17. It names one FDA 510(k) clearance: K060774. Product: Access Folate Calibrators, Part Number: A14207. The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) or red blood cells using the Access Immunoassay Systems. The serum folate level is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency. The Access Folate Calibrators are intended to calibrate the Access Folate assay using the Access Immunoassay Systems.. Reason: Beckman Coulter is recalling the Access Folate Calibrators because it may produce falsely elevated results that could mask a folate deficiency..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2012-07-24
Terminated
2013-10-17
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Access Folate Calibrators, Part Number: A14207. The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) or red blood cells using the Access Immunoassay Systems. The serum folate level is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency. The Access Folate Calibrators are intended to calibrate the Access Folate assay using the Access Immunoassay Systems.

Reason for recall

Beckman Coulter is recalling the Access Folate Calibrators because it may produce falsely elevated results that could mask a folate deficiency.