Recall Z-0044-2026— FUJIFILM Healthcare Americas Corporation

Class II · Ongoing

Class II FDA medical device recall by FUJIFILM Healthcare Americas Corporation, initiated 2025-09-15. It names one FDA 510(k) clearance: K152294. Product: FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33. Reason: It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting..

Recalling firm
FUJIFILM Healthcare Americas Corporation
Classification
Class II
Status
Ongoing
Initiated
2025-09-15
Firm location
Lexington, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

Reason for recall

It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.