Recall Z-0041-2020— Ohmeda Medical
Class II · Terminated
Class II FDA medical device recall by Ohmeda Medical, initiated 2018-10-05, terminated 2020-06-02. It names one FDA 510(k) clearance: K993407. Product: Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).. Reason: GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris..
- Recalling firm
- Ohmeda Medical
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-10-05
- Terminated
- 2020-06-02
- Firm location
- Laurel, MD, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Reason for recall
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.
Data sourced from openFDA. This site is unofficial and independent of the FDA.