Recall Z-0039-2024— AB Sciex, LLC

Class II · Ongoing

Class II FDA medical device recall by AB Sciex, LLC, initiated 2023-08-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Citrine QTRAP Mass Spectrometer, Part Number 5063685. Reason: In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results..

Recalling firm
AB Sciex, LLC
Classification
Class II
Status
Ongoing
Initiated
2023-08-30
Firm location
Framingham, MA, United States

Product description

Citrine QTRAP Mass Spectrometer, Part Number 5063685

Reason for recall

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.