Recall Z-0039-2018— Teleflex Medical

Class II · Completed

Class II FDA medical device recall by Teleflex Medical, initiated 2014-09-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Pilling Knife Handles: a) REF 352950 b) REF 352953 c) REF 352951 d) REF 352957 e) REF 352952 : Reusable surgical knife handles, which are intended for use with blades which are inserted into a slot in the handle.. Reason: The knife handles slot depth is out of specification, so blades may not fit on the handles properly..

Recalling firm
Teleflex Medical
Classification
Class II
Status
Completed
Initiated
2014-09-26
Firm location
Research Triangle Park, NC, United States

Product description

Pilling Knife Handles: a) REF 352950 b) REF 352953 c) REF 352951 d) REF 352957 e) REF 352952 : Reusable surgical knife handles, which are intended for use with blades which are inserted into a slot in the handle.

Reason for recall

The knife handles slot depth is out of specification, so blades may not fit on the handles properly.

Data sourced from openFDA. This site is unofficial and independent of the FDA.